Product Model |
TRID30-7R (Replaceable Battery). |
Sensor Measurement Range |
-30°C to +60°C (-40°F to +210°F). |
Operating Temperature Range |
-30°C to +60°C (-22°F to +140°F). |
Storage Temperature Range |
-10°C to +40°C (14°F to +104°F). |
Rated Temperature Reading Accuracy |
Better than ±0.5°C (±0.9°F) for measurements from -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F).
Better than ±0.8°C (±1.5°F) for other measurements – typically ±0.5°C (0.9°F).
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration. |
Rated Temperature Reading Resolution |
0.1°C (0.2°F) for measurements -40°C to +50°C (-40°F to +122°F).
0.2°C (0.4°F) for measurements +50°C to +70°C (+122°F to +158°F).
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time |
Typically less than 5 minutes (T90) in moving air (1m/s) in EN12830:1999. |
Recording Capacity |
7,770 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging.
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics. |
Sampling Interval |
Configurable from 1 minute to hourly. |
Logging Start Options |
Push button start (with configurable start delay from 1 minute to 18 hours) or specific date & time. |
Recording Indication |
“REC” symbol on LCD |
Download Time |
Typically less than 5 seconds for full memory (7,770 readings), depending on computer or readout device used. |
Environmental |
IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung. |
Power Source |
3V Li-Mg Battery. |
Battery Life |
TRID30-7R: Typically 1 year of operation.
TRID30-7F: Typically 2+ years of operation.
NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily for no longer than 30 seconds each time. Download data monthly. |
Real Time Clock |
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
Connection Interface |
Interface Cradle |
Software |
LogTag® Analyzer |
Size |
93mm(H) x 54.5mm(W) x 8.6mm(T). |
Weight |
TRID30-7R: 41g.
TRID30-7F: 43g. |
Case Material |
Polycarbonate. |
Reviews